Welcome!

This is the online ePortfolio of Jun Hu, Justin Shek (0842536), and Arthur Wong (0957192), students in the Medical Radiation Sciences [Radiography] program with McMaster University and Mohawk College.

Radiographic imaging is a fundamental component of diagnosis in the clinical setting. However, there are various potential sources of error that can produce images that are unusable for diagnosis. The aim of quality control in the radiographic imaging setting is to minimize errors made because of human or apparatus error.

The content of this ePortfolio pertains to the labs, modules, assignments, and assessments of our quality control course, MEDRADSC 3H03: Quality Control in Radiography. By discussing the expectations, goals, and achievements of the aforementioned material, we hope to show how our understanding and appreciation of quality control practices in a radiographic setting will progress.

We have chosen to create this ePortfolio through an online blog format over other formats for three reasons:

1. Various media formats (text, audio, video, images, etc.) are supported
2. Online hosting makes updates and posts possible from virtually anywhere
3. Updates can be added at the poster's leisure and are not restricted to any one user's computer

Furthermore, by posting directly online it is our hope that information and knowledge in this ePortfolio can be more rapidly shared with the world than through conventional means.

Tuesday, March 6, 2012

Timer Accuracy

While the quality of the X-ray beam is affected predominantly by kilovoltage peak (kVp) selection, the quantity of radiation present in the X-ray beam is primarily affected by selection of both milliamperage and exposure time; this selection is collectively referred to as mAs. Appropriate selection of mAs is one of the most influential ways that a technologist can reduce patient dose. It is critical then, that the exposure timer is properly calibrated and accurately describes for how long radiation is emitted from the X-ray tube.

To test this accuracy, this lab was performed by exposing a digital timer, such that the digital timer would display how long an exposure was made as measured to the millisecond. This procedure was performed on two different X-ray tubes. Exposures were made at 80 kVp, large focal spot, 200 mA, and starting from 5 mAs (i.e exposure time = 25 ms), exposure time was doubled after each of five exposures.

Table 1: Exposure factors

Exposure

kVp

mA

mAs

Exposure Time

1

80 kVp

200 mA

5 mAs

25 ms

2

80 kVp

200 mA

10 mAs

50 ms

3

80 kVp

200 mA

20 mAs

100 ms

4

80 kVp

200 mA

40 mAs

200 ms

5

80 kVp

200 mA

80 mAs

400 ms

Image 1: Digital X-Ray Timer

After each exposure, the digital timer would read out how long the actual exposure was, as seen in image 1 above. The evaluation of the accuracy of exposure time was therefore a qualitative one. Table 2 below shows three types of values: the selected exposure time at the control console, the digital timer readouts for both rooms, and the acceptable limits stated by Safety Code 35 and by the HARP Act.

Table 2: Selected Exposure Time, Digital Timer Readouts, and Government Standards

Exposure

Selected Exposure Time

Readout for

Room 3

Readout for

Room 6

SC 35 Standard (within ± 5%)

HARP Act Standard (within 1/30 of a second or 10%, whichever is greater)

1

25 ms

26 ms

26 ms

23.75 to 26.25 ms

23.75 to 26.25 ms

2

50 ms

51 ms

51 ms

47.5 to 52.5 ms

47.5 to 52.5 ms

3

100 ms

101 ms

101 ms

95 to 100 ms

95 to 100 ms

4

200 ms

201 ms

201 ms

190 to 210 ms

190 to 210 ms

5

400 ms

401 ms

401 ms

380 to 420 ms

380 to 420 ms


As Table 2 shows, the results are well within the acceptable limits enforced by Safety Code 35 and the HARP Act. Note that the standards for the HARP Act are those stipulated for a non-mechanical timer. Consequently it appears that no corrective action is needed.

This lab outlines an interesting point to make concerning quality control tests in general. It would be irresponsible to assume that because of the results of this test, corrective action is necessarily not required. However the test relies on a digital timer which itself must be properly functioning and calibrated for the results of this test to be valid. Hence, before the accuracy of the exposure timer can be evaluated it must be known for certain that the digital timer used to measure is also properly functioning. This concept of using properly calibrated test equipment extends to various other tests, including some of the labs discussed in Cluster 1.

As mentioned above, the accuracy of an exposure timer is crucial to an X-ray technologist because it helps reduce patient dose. If the exposure duration is longer than expected then the patient becomes exposed to ionizing radiation for longer than necessary. Furthermore, if the image should become overexposed while in a film-screen environment it may be diagnostically unusable, requiring a repeat image. Conversely, if the exposure duration is too short in either the film-screen or digital radiography environments, the resulting image will also be diagnostically unusable. In cases of under or overexposure, a repeat image is required and patient dose increases. Ensuring a properly functioning exposure timer allows the technologist to uphold the ALARA principle and provide the best patient care possible.

No comments:

Post a Comment